Copper carbonate
Raman SpectroscopyRaman Spectrum
Spectroscopic analysis data
Material
Copper carbonate
Device
- Name
- inVia Confocal Raman microscope
- Alternative Name
- Renishaw microscope
- Manufacturer
- Renishaw
- Serial Number
- 1X2J00
- Category
- RAMAN
- Links
- External Link External Link
Device Parameters
- Detector
- CCD
- Source
- 532 nm
- Location
- Lab Raman OPD
- Objective
- 50× Long distance
- Grating
- 1800 Lines/mm
- Spectral Range
- 100-3600 1/cm
- Integration Time
- 10 s
- Laser Power
- 0.29 mW
- Number Of Scans
- 25
Project & Contributors
- Project
- HSG Spectral Database (hsg-spectral-database)
- Description
- The CNR-INO Spectral Database includes data collected by researchers of the CNR-INO Heritage Science group (HSG). Since 2020, HSG CNR-INO has been working on the application of Raman spectroscopy – a...
- Date
- March 8, 2026
- Analysis ID
- #14
- Analyst
- Diego Ivan Quintero Balbas
- Purpose
- Characterization of chemical composition Corrosion identification Material characterization
Download
TXT
CopperCarbonate_532nm_50xL_10s_0.29mW_25scans_100-3600cm-1.txt
Copper carbonate (532 nm) - Raw
SPC
CopperCarbonate_532nm_50xL_10s_0.29mW_25scans_100-3600cm-1.spc
TXT
CopperCarbonate_532nm_50xL_10s_0.29mW_25scans_100-3600cm-1_Baseline_01.txt
Copper carbonate (532 nm) - Processed
SPC
CopperCarbonate_532nm_50xL_10s_0.29mW_25scans_100-3600cm-1_Baseline_01.spc
Note: Data provided for research and educational purposes.